60.60 Standard published Sep 23, 2022
ISO
ISO/TC 201/SC 6 Secondary ion mass spectrometry
International Standard
71.040.40 Chemical analysis
This document specifies a method for determining the maximum count rate for an acceptable limit of divergence from linearity of the intensity scale in single ion counting time-of-flight (TOF) secondary ion mass spectrometers using a test based on isotopic ratios in spectra from poly(tetrafluoroethylene) (PTFE). It also includes a method to correct for intensity nonlinearity arising from intensity lost from a microchannel plate (MCP) or scintillator and photomultiplier followed by a time-to-digital converter (TDC) detection system caused by secondary ions arriving during its dead-time. The correction can increase the intensity range for 95 % linearity by a factor of up to more than 50 so that a higher maximum count rate can be employed for those spectrometers for which the relevant correction formulae have been shown to be valid.
WITHDRAWN
ISO 17862:2013
PUBLISHED
ISO 17862:2022
60.60
Standard published
Sep 23, 2022