prEN 14180

Sterilizers for medical purposes - Low temperature steam and formaldehyde sterilizers - Requirements and testing prEN 14180

Publication date:   Apr 6, 2023

General information

40.60 Close of voting   Jun 29, 2023

CEN

CEN/TC 102 Sterilizers and associated equipment for processing of medical devices

European Norm

11.080.10   Sterilizing equipment

Buying

Draft

Language in which you want to receive the document.

Scope

1.1 This document specifies requirements and tests for LTSF sterilizers, which use a mixture of low temperature steam and formaldehyde as sterilizing agent, and which are working below ambient pressure only.
These sterilizers are primarily used for the sterilization of heat labile medical devices in health care facilities.
1.2 This document specifies minimum requirements:
- for the performance and design of sterilizers intended to deliver an LTSF-process capable of sterilizing medical devices;
- for the equipment and controls of these sterilizers needed for operation, control and monitoring of the sterilization processes, and which can be used for validation of the sterilization process.
1.3 This document specifies further test equipment and test procedures used to verify conformance of the equipment design and performance specified by this document.
1.4 This document does not specify requirements and tests for decontamination systems for use in rooms, enclosures, or environmental spaces.

Related legislation

Legislation related to this standard

2017/745

Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC

Life cycle

PREVIOUSLY

PUBLISHED
EN 14180:2014

NOW

IN_DEVELOPMENT
prEN 14180
40.60 Close of voting
Jun 29, 2023