ISO 20072:2009

Aerosol drug delivery device design verification — Requirements and test methods ISO 20072:2009

Publication date:   Jul 28, 2009

General information

90.93 Standard confirmed   Mar 9, 2020

ISO

ISO/TC 84 Devices for administration of medicinal products and catheters

International Standard

11.040.10   Anaesthetic, respiratory and reanimation equipment

Buying

Published

Language in which you want to receive the document.

Scope

ISO 20072:2009 applies to the design, labelling, instructions for use and testing requirements for hand-held single- and multi-use aerosol drug delivery devices (ADDDs) intended to deliver a metered or pre-metered aerosolized medication to or by means of the human respiratory tract (including nasal, oral, tracheal, bronchial and alveolar sites). This International Standard applies to both refillable and disposable devices intended for personal use.
ISO 20072:2009 is intended for device design verification and not for drug product quality assessment. The objective of this International Standard is to verify, by laboratory (in-vitro) testing, that the ADDD design consistently meets the manufacturer's design specification by satisfying a device functionality profile and system verification test both of which are determined from a risk assessment and evaluated in accordance with the instructions for use.

Life cycle

NOW

PUBLISHED
ISO 20072:2009
90.93 Standard confirmed
Mar 9, 2020