FprEN ISO 15883-1

Washer-disinfectors - Part 1: General requirements, terms and definitions and tests (ISO/FDIS 15883-1:2024) FprEN ISO 15883-1

Publication date:   Mar 21, 2024

General information

60.55 Ratification completed (DOR)   Jul 12, 2024

CEN

CEN/TC 102 Sterilizers and associated equipment for processing of medical devices

European Norm

11.080.10   Sterilizing equipment

Buying

Draft

Language in which you want to receive the document.

Scope

This document specifies general performance requirements for washer-disinfectors (WD) and washer-disinfector accessories that are intended to be used for cleaning and disinfection of reusable medical devices. It specifies performance requirements for cleaning and disinfection as well as for the accessories that can be required to achieve the necessary performance. The methods and instrumentation required for validation, routine control and monitoring and requalification, periodically and after essential repairs, are also specified.
NOTE 1        The requirements can be applied to WD intended for use with other articles used in the context of medical, dental, pharmaceutical and veterinary practice.
The requirements for WD intended to process specific loads are specified in ISO 15883-2, ISO 15883-3, ISO 15883-4, ISO 15883-6 and ISO 15883-7. For WD intended to process loads of two or more different types, the requirements of the applicable parts of ISO 15883-2, ISO 15883-3, ISO 15883-4, ISO 15883-6 and ISO 15883-7 apply.
This document does not specify requirements intended for machines for use for laundry or general catering purposes.
This document does not include requirements for machines which are intended to sterilize the load, or which are designated as “sterilizers” and addressed in other standards.
The specified performance requirements of this document do not ensure the inactivation or removal of the causative agent(s) (prion protein) of transmissible spongiform encephalopathies.
NOTE 2        Chemicals in some cleaning agents and disinfectants can react with prion protein in a manner that can inhibit its removal or inactivation.  If the presence of prion protein is considered a possibility, then this can influence the choice of cleaning agent and disinfectant.
NOTE 3        This document can be used by prospective purchasers and manufacturers as the basis of agreement on the specification of a WD. The test methods for demonstration of conformity with the requirements of this document can also be employed by users to demonstrate continued conformity of the installed WD throughout its service life. Guidance on a routine test programme is given in Annex A.

Related legislation

Legislation related to this standard

2017/745

Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC

Life cycle

PREVIOUSLY

PUBLISHED
EN ISO 15883-1:2009

PUBLISHED
EN ISO 15883-1:2009/A1:2014

NOW

IN_DEVELOPMENT
FprEN ISO 15883-1
60.55 Ratification completed (DOR)
Jul 12, 2024