60.60 Standard published Jul 28, 2021
CEN
CEN/TC 102 Sterilizers and associated equipment for processing of medical devices
European Norm
11.080.01 Sterilization and disinfection in general
Published
1.1 This document specifies the requirements for a test method to be utilized to establish or confirm a reduced incubation time (RIT) that is shorter than the 7-day reference incubation time specified in 7.3.2 of ISO 11138-1:2017 for biological indicators used to monitor moist heat sterilization processes or ethylene oxide (EO) sterilization processes.
NOTE: For biological indicators used for EO sterilization, the stated RIT is applicable to 100 % EO processes or processes that use EO blends, regardless of the product load.
1.2 This document is applicable to manufacturers of biological indicators (BIs) and to end users of BIs who intend to, if required by their quality system, establish, validate or confirm a RIT.
1.3 This document does not apply to biological indicators used to monitor dry heat, low temperature steam formaldehyde (LTSF) or vaporized hydrogen peroxide (VH2O2) sterilization processes.
NOTE The method described in this document to establish a RIT for biological indicators used to monitor moist heat or EO sterilization processes has been used extensively for many years. However, there is limited experience in use of this method to establish a RIT for biological indicators used to monitor dry heat, low temperature steam formaldehyde or vaporized hydrogen peroxide sterilization processes. This document, therefore, does not include these sterilization processes.
PUBLISHED
EN ISO 11138-8:2021
60.60
Standard published
Jul 28, 2021