ISO/WD 25289

Microphysiological systems and Organ-on-Chip — Key components and minimum requirements for the model of bronchial microphysiological systems ISO/WD 25289

General information

20.60   Close of comment period   Feb 12, 2026

ISO

ISO/TC 276/SC 2

International Standard

Scope

This document specifies (i) characterization parameters and methods used to qualify a bronchial air-liquid interface (ALI) microphysiological system (MPS) for a declared context of use (CoU), and (ii) minimum verification/quality control checks (e.g., lot/batch verification) to confirm that manufactured MPS components continue to meet the qualified performance. It does not establish a full quality management system (QMS) or manufacturing process validation requirements; sampling plans and test frequencies are to be defined by the manufacturer/user and reported. This document covers verification of the bronchial MPS platform. It does not standardize specific experimental test methods. This document provides performance-based guidance for bronchial airway MPS/Organ-on-chip (OoC) with microfluidic control using ALI, where dynamic microenvironmental parameters (e.g., flow/shear, cyclic strain) and multi-compartment architectures with integrated sensing introduce additional sources of variability. Conventional static ALI cultures conducted in standard multi-well plates without a microengineered platform are outside the scope of this document.

Life cycle

NOW

IN_DEVELOPMENT
ISO/WD 25289
20.60 Close of comment period
Feb 12, 2026




Access Genorma App Everywhere