ISO/ASTM 51205:2002

Practice for use of a ceric-cerous sulfate dosimetry system

Publication date:   Apr 18, 2002

95.99 Withdrawal of Standard   Jul 7, 2009

General information

95.99 Withdrawal of Standard   Jul 7, 2009

ISO/ASTM

ISO/TC 85 Nuclear energy, nuclear technologies, and radiological protection

International Standard

17.240   Radiation measurements

Buying

  Withdrawn

PDF - €76.23

  English  



Buy

Scope

ISO/ASTM 51205 covers the preparation, testing and procedure for using the ceric-cerous sulfate dosimetry system to measure absorbed doses in water when exposed to ionizing radiation. The system consists of a dosimeter and appropriate analytical instrumentation. For simplicity, the system is referred to as the ceric-cerous system. It is classified as a reference standard dosimetry system. Ceric-cerous dosimeters are also used as transfer standard dosimeters or routine dosimeters. ISO/ASTM 51205 describes both the spectrophotometric and the potentiometric read-out procedures for the ceric-cerous systems.
The practice described applies only to gamma rays, X rays and high energy electrons and is applicable provided the following are satisfied:
the absorbed-dose range is between 5 times 102 and 5 times 10 4 Gy;
the absorbed-dose rate is less than 106 Gy/s;
for radionuclide gamma-ray sources, the initial photon energy is greater than 0,6 MeV;
for bremsstrahlung photons, the initial energy of the electrons used to produce the bremsstrahlung photons is equal to or greater than 2 MeV;
for electron beams, the initial electron energy is greater than 8 MeV.

Life cycle

PREVIOUSLY

WITHDRAWN
ISO 15555:1998

NOW

WITHDRAWN
ISO/ASTM 51205:2002
95.99 Withdrawal of Standard
Jul 7, 2009

REVISED BY

WITHDRAWN
ISO/ASTM 51205:2009