ISO 7206-13:2016

Implants for surgery — Partial and total hip joint prostheses — Part 13: Determination of resistance to torque of head fixation of stemmed femoral components ISO 7206-13:2016

Publication date:   Jun 27, 2016

General information

90.93 Standard confirmed   Sep 3, 2021

ISO

ISO/TC 150/SC 4 Bone and joint replacements

International Standard

11.040.40   Implants for surgery, prosthetics and orthotics

Buying

Published

Language in which you want to receive the document.

Scope

ISO 7206-13:2016 describes a method of determining the torque required, under specified laboratory conditions, to loosen the fixation of the head of hip joint prostheses in which the head is not intended to rotate relative to the neck. It applies to the femoral component of total or partial hip joint replacements in which the head and neck/stem (in the following referred to as cone) are secured together by a locking conical taper or any other means and in which the head and cone are separate components, and which are made of metallic or non-metallic materials.
It does not cover methods of examining the test specimens; these should be agreed between the test laboratory and the party submitting the specimen for test.

Life cycle

NOW

PUBLISHED
ISO 7206-13:2016
90.93 Standard confirmed
Sep 3, 2021

CORRIGENDA / AMENDMENTS

PUBLISHED
ISO 7206-13:2016/Amd 1:2022