ISO 21501-4:2018

Determination of particle size distribution — Single particle light interaction methods — Part 4: Light scattering airborne particle counter for clean spaces ISO 21501-4:2018

Publication date:   May 18, 2018

General information

90.92   Standard to be revised   May 10, 2026

ISO

ISO/TC 24/SC 4 Particle characterization

International Standard

19.120   Particle size analysis. Sieving

Buying

Published

Language in which you want to receive the document.

Scope

This document describes a calibration and verification method for a light scattering airborne particle counter (LSAPC), which is used to measure the size distribution and particle number concentration of particles suspended in air. The light scattering method described in this document is based on single particle measurements. The typical size range of particles measured by this method is between 0,1 μm and 10 μm in particle size.
Instruments that conform to this document are used for the classification of air cleanliness in cleanrooms and associated controlled environments in accordance with ISO 14644‑1 and ISO 14644‑2, as well as the measurement of number and size distribution of particles in various environments.
The following parameters are within the scope of this document:
— size setting error;
— counting efficiency;
— size resolution;
— false count;
— maximum particle number concentration;
— sampling flow rate error;
— sampling time error;
— response rate;
— calibration interval;
— reporting results from test and calibration.

Life cycle

PREVIOUSLY

WITHDRAWN
ISO 21501-4:2007

NOW

PUBLISHED
ISO 21501-4:2018
90.92 Standard to be revised
May 10, 2026

CORRIGENDA / AMENDMENTS

PUBLISHED
ISO 21501-4:2018/Amd 1:2023

REVISED BY

IN_DEVELOPMENT
ISO/AWI 21501-4




Access Genorma App Everywhere