ISO 14243-3:2014

Implants for surgery — Wear of total knee-joint prostheses — Part 3: Loading and displacement parameters for wear-testing machines with displacement control and corresponding environmental conditions for test ISO 14243-3:2014

Publication date:   Oct 13, 2014

General information

90.92 Standard to be revised   Oct 11, 2024

ISO

ISO/TC 150/SC 4 Bone and joint replacements

International Standard

11.040.40   Implants for surgery, prosthetics and orthotics

Buying

Published

Language in which you want to receive the document.

Scope

ISO 14243-3:2014 describes a test method that specifies flexion/extension relative angular movement between articulating components, the pattern of the applied force, speed and duration of testing, sample configuration and test environment to be used for the wear testing of total knee-joint prostheses in wear-testing machines having axial load control, flexion/extension angular motion control, AP displacement control and tibial rotation control.
The kinematics of this part of ISO 14243-3:2014 may not be applicable to knee designs with a high degree of constraint, which could result in damage to the articulating components in the early stages of the test that would not be representative of clinical service.