ISO 11810-2:2007

Lasers and laser-related equipment — Test method and classification for the laser-resistance of surgical drapes and/or patient-protective covers — Part 2: Secondary ignition ISO 11810-2:2007

Publication date:   Apr 26, 2007

95.99 Withdrawal of Standard   Dec 14, 2015

General information

95.99 Withdrawal of Standard   Dec 14, 2015

ISO

ISO/TC 172/SC 9 Laser and electro-optical systems

International Standard

11.040.30   Surgical instruments and materials | 31.260   Optoelectronics. Laser equipment | 13.340.99   Other protective equipment

Buying

Withdrawn

Language in which you want to receive the document.

Scope

ISO 11810-2:2007 is applicable to disposable and re-usable, as well as woven and non-woven materials used as surgical drapes and/or patient protective covers which claim to be laser-resistant.
The purpose of ISO 11810-2:2007 is to provide a standardized method for testing and classifying surgical drapes and/or patient protective covers with respect to laser-induced hazards. An appropriate classification system is given. It is not the purpose of ISO 11810-2:2007 to serve as a general fire safety specification. ISO 11810-2:2007 is limited to testing the secondary ignition of materials that are rated I1 or I2 from ISO 11810-1.
All materials reflect portions of the beam and it is necessary for the user to decide whether specular reflection may be a hazard. This measurement, however, is not covered in ISO 11810-2:2007.