IEC 60806:2022 ED2

Determination of the maximum symmetrical radiation field of X-ray tube assemblies and X-ray source assemblies for medical diagnosis IEC 60806:2022 ED2

Publication date:   Nov 29, 2022

General information

60.60   Standard published   Nov 29, 2022

IEC

TC 62/SC 62B

International Standard

11.040.50   Radiographic equipment

Buying

Published

Language in which you want to receive the document.

Scope

IEC 60806:2022 is applicable to X-RAY SOURCE ASSEMBLIES and X-RAY TUBE ASSEMBLIES.
This document specifies a method for the determination of the greatest geometrically symmetrical RADIATION FIELD at a specified distance from the FOCAL SPOT for which the percentage AIR KERMA RATE along the major axes of the RADIATION FIELD does not fall below a permitted value.
In case multiple FOCAL SPOTS are not super-imposed, each FOCAL SPOT has its own REFERENCE AXIS. Then the maximum RADIATION FIELD can be given for each FOCAL SPOT separately.
IEC 60806:2022 cancels and replaces the first edition published in 1984. This edition constitutes a technical revision.
This edition includes the following significant technical change with respect to the previous edition:
a) addition of solid state detectors as they have become more common since the first edition of 1984.

Life cycle

PREVIOUSLY

WITHDRAWN
IEC 60806:1984 ED1

NOW

PUBLISHED
IEC 60806:2022 ED2
60.60 Standard published
Nov 29, 2022