IEC 60336:2020 ED5

Medical electrical equipment - X-ray tube assemblies for medical diagnosis - Focal spot dimensions and related characteristics IEC 60336:2020 ED5

Publication date:   Dec 17, 2020

General information

60.60 Standard published   Dec 17, 2020

IEC

TC 62/SC 62B

International Standard

11.040.50   Radiographic equipment

Buying

Published

Language in which you want to receive the document.

Scope

IEC 60336:2020 applies to FOCAL SPOTS in medical diagnostic X-RAY TUBE ASSEMBLIES for medical use, operating at X-RAY TUBE VOLTAGES up to and including 150 kV.
This document describes the test methods employing digital detectors for determining:
a) FOCAL SPOT dimensions in terms of NOMINAL FOCAL SPOT VALUES, ranging from 0,1 to 3,0;
b) LINE SPREAD FUNCTIONS;
c) one-dimensional MODULATION TRANSFER FUNCTIONS;
d) FOCAL SPOT PINHOLE RADIOGRAMS,
and the means for indicating compliance.
In informative annexes, STAR PATTERN imaging and BLOOMING VALUE are described.
IEC 60336:2020 cancels and replaces the fourth edition published in 2005. This edition constitutes a technical revision.
The significant changes of this fifth edition with respect to the previous edition are detailed in Clause E.6. These changes are:
a) introduction of digital detectors and discretization errors;
b) fewer normative requirements;
c) support for both SLIT CAMERA and PINHOLE CAMERA;
d) reintroduction of distorted (skewed) FOCAL SPOT;
e) keeping of STAR PATTERNS and BLOOMING VALUE as informative.
The contents of the corrigendum of June 2022 have been included in this copy.

Life cycle

PREVIOUSLY

WITHDRAWN
IEC 60336:2005 ED4

WITHDRAWN
IEC 60336:2005/COR1:2006 ED4

NOW

PUBLISHED
IEC 60336:2020 ED5
60.60 Standard published
Dec 17, 2020

CORRIGENDA / AMENDMENTS

PUBLISHED
IEC 60336:2020/COR1:2022 ED5