99.60 Withdrawal effective Apr 1, 2011
CEN
CEN/TC 215 Respiratory and anaesthetic equipment
European Norm
11.040.10 Anaesthetic, respiratory and reanimation equipment
Withdrawn
For the purposes of ISO 9919:2005, the Scope given in IEC 60601-1:1988 (Clause 1) applies, except as follows.
Amendment (add at the end of 1.1):
ISO 9919:2005 specifies particular requirements for the basic safety and essential performance of pulse oximeter equipment intended for use on humans. This includes any part necessary for normal use, e.g. the pulse oximeter monitor, pulse oximeter probe, probe cable extender.
These requirements also apply to pulse oximeter equipment, including pulse oximeter monitors, pulse oximeter probes and probe cable extenders, that has been reprocessed.
The intended use of pulse oximeter equipment includes, but is not limited to, the estimation of arterial oxygen haemoglobin saturation and pulse rate on patients in healthcare institutions as well as on patients in home care.
ISO 9919:2005 is not applicable to pulse oximeter equipment intended for use in laboratory research applications nor to oximeters that requires a blood sample from the patient.
ISO 9919:2005 is not applicable to pulse oximeter equipment solely intended for fetal use.
ISO 9919:2005 is not applicable to remote or slave (secondary) devices that display SpO2 values that are located outside of the patient environment.
The requirements of ISO 9919:2005 which replace or modify requirements of IEC 60601-1:1988 and its Amendments 1 (1991) and 2 (1995) are intended to take precedence over the corresponding general requirements.
Legislation related to this standard
Directive 2007/47/EC of the European Parliament and of the council of 5 September 2007 amending Council Directive 90/385/EEC on the approximation of the laws of the Member States relating to active implantable medical devices, Council Directive 93/42/EEC concerning medical devices and Directive 98/8/EC concerning the placing of biocidal products on the market
WITHDRAWN
EN ISO 9919:2005
WITHDRAWN
EN ISO 9919:2009
99.60
Withdrawal effective
Apr 1, 2011
WITHDRAWN
EN ISO 80601-2-61:2011