EN ISO 13163:2019

Water quality - Lead-210 - Test method using liquid scintillation counting (ISO 13163:2013) EN ISO 13163:2019

Publication date:   Jul 15, 2019

General information

99.60 Withdrawal effective   Mar 2, 2022

CEN

CEN/TC 230 Water analysis

European Norm

13.060.60   Examination of physical properties of water | 17.240   Radiation measurements

Buying

Withdrawn

Language in which you want to receive the document.

Scope

ISO 13163 specifies the determination of lead-210 (210Pb) activity concentration in samples of all types of water using liquid scintillation counting (LSC). For raw and drinking water, the sample should be degassed in order to minimize the ingrowth of 210Pb from radon-222 (222Rn).
Using currently available liquid scintillation counters, this test method can measure the 210Pb activity concentrations in the range of less than 20 mBq⋅l-1 to 50 mBq⋅l-1. These values can be achieved with a counting time between 180 min and 720 min for a sample volume from 0,5 l to 1,5 l.
Higher 210Pb activity concentrations can be measured by either diluting the sample or using smaller sample aliquots or both.
It is the laboratory's responsibility to ensure the suitability of this test method for the water samples tested.

Life cycle

NOW

WITHDRAWN
EN ISO 13163:2019
99.60 Withdrawal effective
Mar 2, 2022

REVISED BY

PUBLISHED
EN ISO 13163:2022

Relations

Adopted from ISO 13163:2013 IDENTICAL