60.60 Standard published May 20, 2026
CEN
CEN/TC 206 Biological and clinical evaluation of medical devices
European Norm
11.100.20 Biological evaluation of medical devices
This document specifies allowable limits (AL) for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in EO-sterilized medical devices, procedures for the measurement of EO and ECH, and methods for determining conformity so that devices can be released. Additional background, including guidance and a flowchart showing how this document is applied, are also included in Annexes A, B, C, D, E, F, G, H, I, J and K.
EO-sterilized devices or components that have neither direct nor indirect body or user contact (e.g. in vitro diagnostic devices) are out of scope of this document. This document does not apply to devices that have been demonstrated to not absorb or retain EO or its degradation product ECH, such as medical devices made exclusively of metal alloys and glass, see Clause C.5[228].
NOTE This document does not specify limits for ethylene glycol (EG). No device limits are specified for EG because the risk assessment in Annex F indicates that calculated allowable levels are higher than those likely to occur in a medical device.
Legislation related to this standard
Regulation (EU) 2017/745 of the European Parliament and of the Council of 5 April 2017 on medical devices, amending Directive 2001/83/EC, Regulation (EC) No 178/2002 and Regulation (EC) No 1223/2009 and repealing Council Directives 90/385/EEC and 93/42/EEC
WITHDRAWN
EN ISO 10993-7:2008
WITHDRAWN
EN ISO 10993-7:2008/AC:2009
WITHDRAWN
EN ISO 10993-7:2008/A1:2022
PUBLISHED
EN ISO 10993-7:2026
60.60
Standard published
May 20, 2026