EN 61010-031:2015/A11:2021

Safety requirements for electrical equipment for measurement, control and laboratory use - Part 031: Safety requirements for hand-held and hand-manipulated probe assemblies for electrical test and measurement EN 61010-031:2015/A11:2021

Publication date:   Mar 23, 2022

General information

60.60 Standard published   Nov 12, 2021

CENELEC

CLC/TC 66X Safety of measuring, control, and laboratory equipment

European Norm

19.080   Electrical and electronic testing

Buying

Published

Language in which you want to receive the document.

Scope

1.1.1 Probe assemblies included in scope

This part of IEC 61010 specifies safety requirements for hand-held and hand-manipulated probe assemblies of the types described below, and their related accessories. These probe assemblies are for direct electrical connection between a part and electrical test and measurement equipment. They may be fixed to the
equipment or be detachable accessories for the equipment.

a) Type A: low-voltage and high-voltage, non-attenuating probe assemblies. Non-attenuating probe assemblies that are RATED for direct connection to voltages exceeding 30 V r.m.s., 42,4 V peak, or 60 V d.c., but not exceeding 63 kV. They do not incorporate components which are intended to provide a voltage divider function or a signal conditioning function, but they may contain non-attenuating components such as
fuses (see Figure 1).

b) Type B: high-voltage attenuating or divider probe assemblies. Attenuating or divider probe assemblies that are RATED for direct connection to secondary voltages exceeding 1 kV r.m.s or 1,5 kV d.c. but not exceeding 63 kV r.m.s or d.c. The divider function may be carried out wholly within the probe assembely, or partly within the test or measurement equipment to be used with the probe assembly (see Figure 2).

c) Type C: low-voltage attenuating or divider probe assemblies. Attenuating or divider probe assemblies for direct connection to voltages not exceeding 1 kV r.m.s. or 1,5 kV d.c. The signal conditioning function may be carried out wholly within the probe assembly, or partly within the test or measurement equipment intended to be used with the probe assembly (see Figure 3).

d) Type D: low-voltage attenuating, non-attenuating or other signal conditioning probe assemblies, that are RATED for direct connection only to voltages not exceeding 30 V r.m.s., or 42,4 V peak, or 60 V d.c., and are suitable for currents exceeding 8 A (see Figure 4).

1.1.2 Probe assemblies excluded from scope

This standard does not apply to current sensors within the scope of IEC 61010-2-032 (hand-held and hand-manipulated current sensors), but may apply to their input measuring circuit leads and accessories.

Related legislation

Legislation related to this standard

2014/35/EU

Directive 2014/35/EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of electrical equipment designed for use within certain voltage limits

Life cycle

PREVIOUSLY

PUBLISHED
EN 61010-031:2015/A1:2021

NOW

PUBLISHED
EN 61010-031:2015/A11:2021
60.60 Standard published
Nov 12, 2021

REVISED BY

PUBLISHED
EN IEC 61010-031:2023