40.00 DIS registered Nov 21, 2024
ISO
ISO/TC 201/SC 7 Electron spectroscopies
International Standard
71.040.40 Chemical analysis
ISO 24237:2005 specifies a method for evaluating the repeatability and constancy of the intensity scale of X-ray photoelectron spectrometers, for general analytical purposes, using unmonochromated Al or Mg X-rays or monochromated Al X-rays. It is only applicable to instruments that incorporate an ion gun for sputter cleaning. It is not intended to be a calibration of the intensity/energy response function. That calibration may be made by the instrument manufacturer or other organization. The present procedure provides data to evaluate and confirm the accuracy with which the intensity/energy response function remains constant with instrument usage. Guidance is given on some of the instrument settings that may affect this constancy.
PUBLISHED
ISO 24237:2005
IN_DEVELOPMENT
ISO/DIS 24237
40.00
DIS registered
Nov 21, 2024