ISO 16063-17:2016

Methods for the calibration of vibration and shock transducers — Part 17: Primary calibration by centrifuge ISO 16063-17:2016

Publication date:   May 30, 2016

General information

90.93 Standard confirmed   Dec 17, 2021

ISO

ISO/TC 108 Mechanical vibration, shock and condition monitoring

International Standard

17.160   Vibrations, shock and vibration measurements

Buying

Published

Language in which you want to receive the document.

Scope

ISO 16063 comprises a series of documents dealing with methods for the calibration of vibration and shock transducers.
ISO 16063-17:2016 lays down detailed specifications for the instrumentation and procedure to be used for primary calibration of accelerometers using centrifuge calibration.
ISO 16063-17:2016 is applicable to rectilinear accelerometers with zero-frequency response, mainly of the strain gauge or piezoresistive type, and to primary standard and working transducers.
It is applicable for a calibration range from 10 m/s2 to 20 000 m/s2 (higher accelerations possible) at 0 Hz.
The limits of uncertainty applicable are ±1 % of reading.

Life cycle

PREVIOUSLY

WITHDRAWN
ISO 5347-7:1993

NOW

PUBLISHED
ISO 16063-17:2016
90.93 Standard confirmed
Dec 17, 2021