IEC 60645-6:2022 ED2

Electroacoustics - Audiometric equipment - Part 6: Instruments for the measurement of otoacoustic emissions IEC 60645-6:2022 ED2

Publication date:   Mar 10, 2022

General information

60.60 Standard published   Mar 10, 2022

IEC

TC 29

International Standard

17.140.50   Electroacoustics

Buying

Published

Language in which you want to receive the document.

Scope

IEC 60645-6:2022 is available as IEC 60645-6:2022 RLV which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.

IEC 606045-6:2022 applies to instruments designed primarily for the measurement of otoacoustic emissions in the human external auditory meatus evoked by acoustic probe stimuli. This document defines the characteristics to be specified by the manufacturer, specifies minimum mandatory functions for two types of instruments and provides performance specifications applicable to both instrument types. This document describes methods to be used to demonstrate conformance with the specifications in this document and guidance on methods for periodic calibration. The purpose of this document is to ensure that measurements made under comparable test conditions with different instruments complying with this document will be consistent. Instruments can provide a measurement function not specifically within the scope of this document and still comply with the relevant requirements of this document for the functions that are within the scope. This document is not intended to restrict development or incorporation of new features, nor to discourage innovative approaches. IEC 606045-6:2022 cancels and replaces the first edition published in 2009. This edition constitutes a technical revision.
This edition includes the following significant technical changes with respect to the previous edition:
a) the nominal test frequency used in DPOAE is now defined as the higher of the two frequencies, f2;
b) the permitted deviation of the stimulus signal for TEOAE has been specified;
c) the frequency range for DPOAE stimulus signals has been redefined,
d) the stimulus level requirements for TEOAE have been redefined;
e) the stimulus level requirements for DPOAE have been redefined;
f) the harmonic distortion requirements for DPOAE have been redefined;
g) a minimum measurement range for DPOAE has been added.

Life cycle

PREVIOUSLY

WITHDRAWN
IEC 60645-6:2009 ED1

NOW

PUBLISHED
IEC 60645-6:2022 ED2
60.60 Standard published
Mar 10, 2022