60.60 Standard published Sep 10, 2021
CENELEC
CLC/TC 66X Safety of measuring, control, and laboratory equipment
European Norm
19.080 Electrical and electronic testing | 71.040.20 Laboratory ware and related apparatus
Published
<span style="color:red">NEW!</span>IEC 61010-2-061:2018 is available as <a href="https://webstore.iec.ch/publication/63899">IEC 61010-2-061:2018 RLV</a> which contains the International Standard and its Redline version, showing all changes of the technical content compared to the previous edition.</br></br>IEC 61010-2-061:2018 applies to electrically powered laboratory atomic spectrometers with thermal atomization. It has the status of a group safety publication in accordance with IEC Guide 104. This fourth edition cancels and replaces the third edition published in 2015. This edition constitutes a technical revision. This edition includes the following significant technical changes with respect to the previous edition:
- adaptation of changes introduced by Amendment 1 of IEC 61010-1;
- added tolerance for stability of AC voltage test equipment to Clause 6;
- added requirement for interlock systems containing electric/electronic or programmable components to Clause 15.
Legislation related to this standard
Directive 2014/35/EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of electrical equipment designed for use within certain voltage limits
Harmonized/Supporting
WITHDRAWN
EN 61010-2-061:2015
PUBLISHED
EN IEC 61010-2-061:2021
60.60
Standard published
Sep 10, 2021
PUBLISHED
EN IEC 61010-2-061:2021/A11:2021